Skip to main content

Research Repository

Advanced Search

Variability of automated carotid intima-media thickness measurements by novice operators.

Nichols, S.; Milner, M.; Meijer, R.; Carroll, S.; Ingle, L.

Authors

M. Milner

R. Meijer

S. Carroll

L. Ingle



Abstract

Carotid intima-media thickness (C-IMT) measurements provide a non-invasive assessment of subclinical atherosclerosis. The aim of the study was to assess the inter- and intra-observer variability of automated C-IMT measurements undertaken by two novice operators using the Panasonic CardioHealth Station. Participants were free from cardio-metabolic disease, and each underwent serial bilateral C-IMT ultrasound measurements. Immediate interoperator measurement variability was calculated by comparing initial measurements taken by two operators. Immediate retest variability was calculated from two consecutive measurements and longer term variability was assessed by conducting a further scan 1 week later. Fifty apparently healthy participants (n = 20 females), aged 26.2 ± 5.0 years, were recruited. Operator 1 recorded a median (interquartile range) right and left-sided C-IMT of 0.471 mm (0.072 mm) and 0.462 mm (0.047 mm). Female's right and left C-IMT were 0.442 mm (0.049 mm) and 0.451 mm (0.063 mm), respectively. The limits of agreement (LoA) for immediate interoperator variability were −0.063 to 0.056 mm (mean bias −0.003 mm). Operator 1's immediate retest intra-operator LoA were −0.057 to 0.046 mm (mean bias was −0.005 mm). One-week LoA were −0.057 to 0.050 mm (mean bias −0.003 mm). Operator 2 recorded median right and left-sided C-IMT of 0.467 mm (0.089 mm) and 0.458 mm (0.046 mm) for males, respectively, whilst female measurements were 0.441 mm (0.052 mm) and 0.444 mm (0.054 mm), respectively. Operator 2's intra-operator immediate retest LoA were −0.056 to 0.056 (mean bias <−0.001 mm). Intra-operator LoA at 1 week were −0.052 to 0.068 mm (mean bias 0.008 mm). Novice operators produce acceptable short-term and 1-week inter- and intra-operator C-IMT measurement variability in healthy, young to middle-aged adults using the Panasonic CardioHealth Station.

Citation

NICHOLS, S., MILNER, M., MEIJER, R., CARROLL, S. and INGLE, L. 2016. Variability of automated carotid intima-media thickness measurements by novice operators. Clinical physiology and functional imaging [online], 36(1), pages 25-32. Available from: https://doi.org/10.1111/cpf.12189

Journal Article Type Article
Acceptance Date Aug 15, 2014
Online Publication Date Sep 12, 2014
Publication Date Jan 31, 2016
Deposit Date Dec 13, 2023
Publicly Available Date Dec 13, 2023
Journal Clinical physiology and functional imaging
Print ISSN 1475-0961
Electronic ISSN 1475-097X
Publisher Wiley
Peer Reviewed Peer Reviewed
Volume 36
Issue 1
Pages 25-32
DOI https://doi.org/10.1111/cpf.12189
Keywords Cartoid intima-media thickness; Ultrasound; Atherosclerosis; Health care workers; Health technology
Public URL https://rgu-repository.worktribe.com/output/2079790